Amplified output signal o Temperature limits:
-40 +350C for type T
-40 +750C for type J
-40 +900C for type E
-40 +1000C for type K (see table for sheath diameters)
Tolerance:
IEC 584.2 STANDARD class 1:
or type T:
0, 5 C (- 40 + 125 C)
0.004 [t] (t > + 125 C)
or type J, E, K:
1.5 C (- 40 + 375 C)
0.004 [t] (t > + 375 C)
Mineral insulated (MgO) thermocouple element and continuous metal sheath
Application versatility